Skip to main content

Table 4 Comparison between agreed and disagreed groups

From: A novel method for diaphragm-based electrode belt position of electrical impedance tomography by ultrasound

Parameters

Agreed group

(n = 23)

Disagreed group

(n = 27)

P-value

Age (yrs)

55 ± 14

62 ± 14

0.085

Heart rate (bpm)

81 ± 16

89 ± 14

0.09

MAP (mmHg)

94 ± 14

90 ± 14

0.330

APACHE-II

12 ± 5

14 ± 6

0.266

BMI

23 ± 2

26 ± 3

0.002*

PEEP (cmH2O)

5.5 ± 1

5.4 ± 1

0.955

Oxygenation index

334 ± 135

294 ± 115

0.231

Tidal volume (ml)

467 ± 98

423 ± 73

0.092

Glob-TVDiaphragm-based EIT-belt plane (AU)

3298 ± 1139

2898 ± 1059

0.263

Glob-TV5th ICS plane (AU)

3214 ± 1120

2246 ± 968

0.005*

TV/VT ratio Diaphragm-based EIT-belt plane

7.3 ± 2.9

7.1 ± 2.9

0.884

TV/VT ratio 5th ICS plane

7.1 ± 2.9

5.5 ± 2.7

0.04*

  1. PEEP positive end-expiratory pressure, TV/VT tidal impedance variance/tidal volume, BMI body mass index, AU arbitrary unit
  2. *Agreed group vs. disagreed group, P < 0.05
  3. Diaphragm-based EIT-belt plane vs. 5th ICS plane, P < 0.05